Updated November 5, 2026 · By Andy Leleux, sleep technologist
Why it really matters
Cleaning is usually framed as a hygiene issue, and it is: the manuals point out that germs can grow in your equipment. But there is a mechanical reason too. A dirty or worn cushion seals less well. It leaks, and a large leak lowers the pressure you actually receive, with no warning at all.
When your CPAP "doesn't work like it used to," check the cushion first, before you start questioning the settings.
Every morning
The easiest time is right when you get up and take the mask off. Wait until evening, and skin oils have had all day to dry on.
The mask cushion
Wash it in warm water (about 86°F/30°C) with a mild liquid detergent, using a soft brush on the small vent holes. Rinse under running water, then let it air-dry out of direct sunlight.
This is the part that sits against your skin all night. Both ResMed and Philips say to wash it every day.
The humidifier tank
Empty it, wipe it with a clean cloth and let it dry out of direct sunlight. Refill it at bedtime with room-temperature water.
That is ResMed's daily instruction. Never use hot water: the air reaching your mask would get too warm.
Your face
Wash your face before you put the mask on, and skip moisturizer on your face and hands.
The Philips manual says exactly that. Oils and creams damage the silicone and keep it from sealing.
Once a week
Pick a set day, ideally in the morning, so everything has all day to dry.
The tubing
Warm water and mild dish soap, rinse well, then let it dry away from sunlight and heat.
No dishwasher and no washing machine: ResMed rules both out for tubing. Hang it up so all the water drains out.
The headgear and frame
Soak for up to ten minutes in warm water with a mild detergent, then rinse by squeezing them under running water.
Don't wring the headgear. Squeeze the water out instead, as ResMed recommends.
The humidifier tank
Warm water and mild dish soap, or 1 part vinegar to 9 parts water at room temperature. The same mix removes white mineral deposits.
Rinse thoroughly with clean water, then let it dry. Nothing hotter than 131°F (55°C).
The outside of the machine
A dry cloth, with the machine unplugged.
Never put the machine in water, and only plug it back in once it is completely dry.
What the manuals say
| Part | How often | Source |
|---|---|---|
| Cushion | Every day | ResMed AirFit N20, Philips DreamWear |
| Headgear and frame | Every week | ResMed AirFit N20 |
| Humidifier tank | Emptied daily, washed weekly | ResMed AirSense 11 |
| Tubing | Every week | ResMed AirSense 11 |
| Filter (AirSense 11) | Replaced every six months, not washable | ResMed AirSense 11 |
Never put a mask back together while it is still damp. The manuals say to let every part dry before you reassemble the mask, and to store it only when it is clean and dry. Put it back together in the evening if you need to.
Not every manual says the same thing. Philips allows the DreamWear mask in the dishwasher once a week. ResMed doesn't list it for the AirFit N20 mask, but does allow it for the AirSense 11 water tank. Schedules also vary from one model to the next.
What never changes: warm water, a mild detergent, a thorough rinse, and drying out of direct sunlight. For everything else, your equipment's manual and your sleep center's instructions take priority over this page. If you are unsure, ask at your next follow-up visit.
Filters follow their own schedule, which depends on the model. See how long a CPAP lasts for the details.
What water goes in the humidifier?
Whatever your tank's manual says. For the ResMed AirSense 11, the standard tank takes distilled water only, while the cleanable tank for the same machine accepts drinking water. If in doubt, use distilled water (or demineralized water with no fragrance or additives).
Tap water leaves mineral deposits in the tank. ResMed says to remove that white residue with diluted vinegar, and to replace a tank that leaks, is cracked, or has turned cloudy or pitted.
If tap water is all you have, say on a trip, a few nights won't hurt. Rinse the tank more often and switch back to the water your manual recommends as soon as you can. See also traveling with CPAP.
What not to do
Five mistakes to avoid, from most to least costly.
Ozone or UV cleaners
The FDA has not authorized any device to clean, disinfect or sterilize a CPAP, and it has received reports of asthma attacks, headaches and shortness of breath after people used ozone devices. Both ozone and UV light break down tubing and masks.
The dishwasher, unless your manual allows it
Some parts can go in, others can't, and tubing never can. A damaged part seals less well: it leaks, and the pressure you receive drops.
Bleach, chlorine, alcohol, alcohol wipes
ResMed and Philips both rule them out. They can damage your equipment or leave fumes behind that you breathe in if you don't rinse well enough.
Antibacterial, moisturizing or scented soaps, scented oils
ResMed rules these out for the machine, tank and tubing: they can damage them and shorten their life. Mild dish soap is enough, and it is the example ResMed gives. Then rinse well.
Direct sunlight and heat
Every manual says to air-dry out of direct sunlight. None of them mentions a hair dryer.
A word about automatic cleaners
You see them everywhere online: devices that promise to disinfect your CPAP in minutes, with no water, using ozone or ultraviolet light.
The FDA is clear: no device is cleared or approved to clean, disinfect or sterilize a CPAP. Only one has been authorized, and for a much narrower purpose: reducing bacteria on certain specific masks and tubing, on top of your regular cleaning, never instead of it.
The main risk is to you, not the machine. The FDA says it has received reports from CPAP users of asthma attacks, headaches and shortness of breath after using ozone devices. It adds that ozone can leak into the room during the cycle, and that levels inside the equipment can stay above safe limits several hours after cleaning ends. And that is the circuit you then breathe through all night.
The same source makes two more points: ozone and UV light can break down tubing and masks, and the FDA has received no data showing that either one cleans the inside of CPAP tubing.
As for manufacturers, ResMed says its limited warranty excludes ozone damage since February 1, 2020, and notes that its manuals call for warm water and a mild detergent only. During the 2021 recall of some of its machines, Philips wrote that ozone may contribute to breakdown of their foam, and that no ozone or UV cleaner is approved for its machines or masks.
In short: a device that isn't authorized for anything it claims to do, that users have linked to breathing symptoms, that wears out your equipment, and whose damage ResMed's warranty won't cover. Warm water and a mild detergent are still what the manuals and the FDA recommend. Some manufacturers also accept diluted vinegar, but check your manual first.
Even with good care, parts wear out
Silicone ages even when you clean it properly. ResMed doesn't give a fixed lifespan for a mask: it depends on how much you use it, how you care for it and how you store it. Instead, the manuals list signs to look for.
Cushion and mask
Cracks, crazing, tears
Or as soon as it stops sealing, even when properly fitted
Humidifier tank
Leaking, cracked, cloudy or pitted
Or a cracked or torn seal
Tubing
Holes, tears, cracks
Check during the weekly wash
Filter (AirSense 11)
Every six months
Sooner if it has holes or is clogged
Four useful tips
- A soft brush for the small holes. ResMed says to pay special attention to the vent holes with a soft-bristle brush. For the inside of the tubing, a long, flexible bottle brush helps remove what water alone leaves behind.
- If your tubing is heated (ClimateLineAir on ResMed machines), it has wiring inside: follow its manual, and never put it in the dishwasher or washing machine.
- Bring your CPAP if you go into the hospital, along with your mask, and tell the staff. A machine with your settings may not be available there.
- Include your CPAP on your list of treatments, at the doctor's office and at the hospital. It isn't a medication, but it is a treatment, and people don't always think to ask.
Warning signs
- A smell that lingers after washing: it is often the tubing, not the mask.
- Black spots in the tubing or tank: stop using the humidifier and get the part replaced.
- A tank that has turned cloudy or pitted, or cracked tubing.
- A cushion that feels sticky, looks dull, or leaves a deep mark in the morning: it is near the end of its life.
- Leaks that start even though nothing has changed in your settings.
In any of these cases, talk to your sleep center. Under the NIHDI (INAMI/RIZIV) convention, the center supplies your mask, filters and tubing, and replaces them free of charge when they are defective. There is one exception: damage caused by misuse, loss or poor maintenance can be billed to you. The center can also swap a mask that doesn't suit you, even if it isn't damaged. Everything is explained on the CPAP convention page. Your center's direct CPAP follow-up number, when we have it, is on its listing.
Need a spare part outside the convention?
If you buy your own equipment, or want a spare cushion or headgear on hand, VivaRespire sells replacement parts.
The author of this site runs VivaRespire, which helps fund this site. FR
Frequently asked questions
Sources
- U.S. Food and Drug Administration, Do You Need a Device That Claims to Clean a CPAP Machine?, updated August 26, 2024. Accessed November 5, 2026. fda.gov
- ResMed, Limited warranty: ozone damage excluded as of February 1, 2020; cleaning with warm water and mild detergent. Accessed November 5, 2026. resmed.com
- ResMed, AirSense 11 User Guide: daily and weekly cleaning, tank water, 1:9 vinegar solution, checking and replacing parts (copy of the official guide). PDF
- ResMed, AirFit N20 User Guide: mask cleaning, service life, visual inspection criteria (copy of the official guide). PDF
- Philips Respironics, DreamWear Nasal Mask, Instructions for Use: daily and weekly washing, dishwasher, products to avoid. documents.philips.com
- Philips, Ozone Cleaner Information, February 8, 2022. philips.co.uk
- NIHDI (INAMI/RIZIV), rehabilitation convention for obstructive sleep apnea syndrome, in force since January 1, 2018, art. 7 § 3 (supplying and replacing equipment) and art. 17 (repair and replacement costs). In French. inami.fgov.be
Go further
Mask still leaking even though you clean it properly?
A leak that persists with a clean cushion in good shape usually comes down to the mask itself or how it fits, not dirt.